94
Programming
Making Measurements
Measure delta, drift and signal-to-noise
To select a measurement, use one of the following STATe commands:
CALC3:DELT:POW:STAT
(delta power)
CALC3:DELT:WAV:STAT
(delta wavelength)
CALC3:DELT:WPOW:STAT
(delta power and wavelength)
CALC3:DRIF:STAT
(drift)
CALC3:SNR:STAT
(signal-to-noise ratios)
CALC3:ASNR:STAT
(signal-to-noise ratio averaging)
If you select a drift measurement, you can additionally select one of the following
additional states:
CALC3:DRIF:DIFF:STAT
(difference)
CALC3:DRIF:MAX:STAT
(maximum drift)
CALC3:DRIF:MIN:STAT
(minimum drift)
CALC3:DRIF:REF:STAT
(drift reference values)
The :CALCulate3:DRIFt:PRESet command turns off the minimum, maximum, dif-
ference, and reference states but leaves the drift state on.
Attempting to turn more than one state on at a time results in a
“–221 Settings Con-
flict”
error.
The *RST and SYSTem:PRESet commands turn all calculations off.
CALCulate3:PRESet turns off any CALCulate3 calculations.
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......