55
Using the Multi-Wavelength Meter
Defining Laser-Line Peaks
2
Press the
THRSHLD
softkey.
3
Press
PX EXC
, and enter the peak excursion value. Use the
softkey to select the
digit that requires editing. Use the
and
softkeys to change the value.
The peak excursion value can range from 1 to 30 dB. The default value is 15 dB.
4
Press
RETURN
.
5
Press
PK THLD
and then enter the peak threshold value.
The peak threshold value can range from 0 to 40 dB. Setting this value to 0 dB en-
sures that only the peak wavelength is identified. The default value is 10 dB.
Pressing the green PRESET key changes the peak excursion and peak threshold val-
ues to their default settings. It also turns wavelength range limiting on. Turning the
Agilent 86120B’s power off and then on does not change these settings.
If too many lines are identified
If the following message is displayed, too many laser lines have been identified:
E15 MAX NUMBER OF SIGNALS FOUND
The maximum number of laser lines that the instrument can measure is 100. If
this message appears, decrease the peak threshold value, increase the peak
excursion value, or decrease the wavelength range of operation with the
WL
LIM
....
START WL
and
STOP WL
functions.
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......