Example 6. Increase a source’s wavelength accuracy
This example program uses the Keysight 86120B to increase the absolute wave-
length accuracy of Keysight 8167A, 8168B, and 8168C Tunable Laser Sources.
Essentially, the Keysight 86120B’s accuracy is transferred to the tunable laser
source. The absolute accuracy of the tunable laser source is increased from <
±
0.1
nm to <
±
0.005 nm which is the Keysight 86120B’s absolute accuracy (at 1550 nm).
In order to run this program, the tunable laser source’s firmware must support the
automatic alignment command, WAVEACT.
The program uses the following measurement algorithm:
Identify and initialize the Keysight 86120B and tunable laser source
Ask user for desired wavelength
Set wavelength of tunable laser source
Turn tunable laser source’s output on
Enter loop
Measure wavelength
Compare wavelength to desired wavelength
Realign tunable laser source’s wavelength
Check if wavelength changed from last pass
Repeat until (delta wavelength < 0.0015 nm
or
wavelength is stable)
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......