153
Programming Commands
Measurement Instructions
DEFault
Current resolution
Examples
:CONF:ARR:POW:WAV DEF MAX
:FETC:ARR:POW:WAV? DEF MIN
:READ:ARR:POW:WAV?
:MEAS:ARR:POW:WAV?
:CONF:SCAL:POW:WAV 1300NM, MAX
:FETC:SCAL:POW:WAV? 1300NM, MIN
:READ:SCAL:POW:WAV? 1300NM
:MEAS:SCAL:POW:WAV? 1300NM
Query Response
The following line is an example of a returned string when
:MEAS:SCAL:POW:WAV? MAX is sent:
+1.5529258E-006
If six laser lines are located and :MEAS:ARR:POW:WAV? is sent, the following
string could be returned. Notice that the first returned number indicates the number
of laser-line values returned in the query.
6,+1.54488100E-006,+1.54648400E-006,+1.54809000E-006,+1.54969900E-
006,+1.55131100E-006,+1.55292600E-006