77
Measurements Applications
Measuring Laser Drift
wavelength and
smallest
power measured. The laser line of
interest may have since drifted to a greater value. Note that the
minimum wavelength and minimum power may not have
occurred simultaneously.
Display shows the
total
drift from the reference since the drift
measurement was started. Values represent the minimum
wavelength and power drift values subtracted from the
maximum drift values.
5
In the List by WL and List by Power displays, use the
and
softkeys to view
the reference values (wavelength and power values of each laser line before the test
was started).
During the measurement, you can change the display mode to Peak WL, List by WL,
List by Power, or Avg WL. When List by WL or List by Power is selected, the signal
list is sorted by reference values and not by the current, maximum, or minimum val-
ues.
To restart the drift measurements, press
RESET
. This resets the reference values.
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......