158
Programming Commands
CALCulate1 Subsystem
DATA?
Queries
uncorrected
frequency-spectrum data of the input laser line.
Syntax
:CALCulate1:DATA?
Attribute Summary
Preset State: not affected
SCPI Compliance: standard
Query Only
Description
The returned values are in squared Watts (linear) units. No amplitude or frequency
correction is applied to the values. To obtain the logarithmic (dB) result, normalize
the returned values by the largest value, then take five times the logarithm of the
returned values.
Be prepared to process a large amount of data when this query is sent. The amount
of data returned depends on the measurement update state of the instrument which
can be set using the :CALCulate1:TRANsform:FREQuency:POINts command or
the resolution argument of an instrument function.
Refer to “Measurement Instruc-
When
NORMAL
measurement update is specified, over 580 kilobytes of data
(34,123 values) can be returned to the computer. When
FAST
measurement update is
specified, over 72 kilobytes of data (4,268 values) can be returned.
The following string is a typical example of the first few returned values:
+4.02001,+6.78001,+6.17001,+4.26001,+4.802453
00E+001,+3.10001,+1.13001,+5.07001,+2.77001,
+3.89001,+3.50001,+7.34649800E-001,+5.64000,
Notice that only measurement values are returned to the computer. There is no first
value that indicates the number of values contained in the string as there is, for
example, with the FETCh, READ, and MEASure commands. Use the :CALCu-
late1:TRANsform:FREQuency:POINTs command to query the number of points the
:CALC1:DATA? returns.
When
NORMAL
measurement mode is selected, the uncorrected frequency domain
data consists of 64K (65,536) values. Only the frequency domain data correspond-
ing to 700-1650 nm wavelength (in vacuum) is returned (34,123 values). In
FAST
measurement mode, the data consists of 8K (8,192) values of which 4,268 values are
returned.
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......