![Keysight Technologies 86120B User Manual Download Page 210](http://html1.mh-extra.com/html/keysight-technologies/86120b/86120b_user-manual_1961930210.webp)
208
Programming Commands
STATus Subsystem
Query Response
When queried, the largest value that can be returned is 65535. This is because the
most-significant register bit cannot be set true.
{OPERation | QUEStionable}[:EVENt]
Queries the contents of the questionable or operation event registers.
Syntax
:STATus:{OPERation | QUEStionable}:EVENt?
Query Response
0 to 32767
Attribute Summary
Preset State: none
*RST State: none
SCPI Compliance: standard
Query Only
Description
The response will be a number from 0 to 32767 indicating which bits are set. Read-
ing the register clears the register.
Example
OUTPUT 720;”:STATUS:OPERATION:EVENT?”
{OPERation | QUEStionable}:NTRansition
Selects bits in the event register which can be set by negative transitions of the cor-
responding bits in the condition register.
Syntax
:STATus:OPERation:NTRansition{?|
<integer>}
<integer>
an integer from 0 to 65535.
Attribute Summary
Preset State: none
*RST State: none
SCPI Compliance: standard
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......