185
Programming Commands
CALCulate3 Subsystem
DRIFt:DIFFerence[:STATe]
Sets the drift calculation to subtract the minimum values measured from the maxi-
mum values measured.
Syntax
:CALCulate3:DRIFt:DIFFerence[:STATe]{?| {ON | OFF | 1 | 0}}
Attribute Summary
Preset State: off
*RST State: off
SCPI Compliance: instrument specific
Description
Use the CALC3:DRIF:PRES command to turn off all the drift states before turning
on this state. The CALC3:DATA? query returns the maximum power and frequency
minus the minimum power and frequency.
Note
Only one STATe command can be turned on at any one time. Attempting to turn
more than one state on at a time results in a “–221 Settings Conflict” error.
Refer to “Measure delta, drift and signal-to-noise” on page 94
for additional
information on selecting measurements.
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......