205
Programming Commands
SENSe Subsystem
Syntax
:SENSe:DATA?
Attribute Summary
Preset State: none
SCPI Compliance: instrument specific
Query Only
Description
Be prepared to process a large amount of data when this query is sent. The amount
of data returned depends on the measurement update state of the instrument which is
set using the resolution argument of an instrument function.
.
When
NORMAL
measurement update is specified, over 2,200 kilobytes of data
(128K values) can be returned to the computer. When
FAST
measurement update is
specified, over 250 kilobytes of data (16K values) can be returned.
The floating point values are scaled from 1.000 to 1.999 (1 + 1023/1024). Ampli-
tude values are not calibrated.
The input laser line(s) generate an interference pattern on the photodetector as a
function of the Michelson interferometer optical path delay. The time domain data is
sampled at uniform optical path delay increments of half the reference laser wave-
length, or 0.316495 microns. When
NORMAL
measurement update is selected, the
first data value is sampled at –20.74 mm optical path delay, and the last value is
sampled at +20.74 mm optical path delay. When FAST measurement update is
selected, the first data value is sampled at +2.59 mm optical path delay, and the last
value is sampled at –2.59 mm optical path delay. The data value that corresponds to
zero optical path delay is approximately, but not exactly, located in the center of the
time domain data.
If your program is aborted or interrupted after sending this query, the
Keysight 86120B continues to process the data but does not place it in the output
buffer. Because of the amount of data processed, the instrument will not respond to
any new commands in its input buffer for 30 or 40 seconds.
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......