45
Using the Multi-Wavelength Meter
Displaying Wavelength and Power
units are later changed, the start and stop wavelength units will change accordingly.
Note that a start wavelength limit in nm will become a stop wavelength limit if THz
or cm
-1
is chosen. See
“To change the units of measure” on page 49
.
The wavelength limit can be useful when laser modulation causes spurious wave-
lengths to be displayed. Reducing the wavelength range to the region of interest
minimizes the number of spurious wavelengths displayed. Also, the graphical dis-
play uses these start and stop wavelength values to plot the power spectrum, whether
the wavelength limit function is on or off.
Preset will turn the wavelength limit on and will set the start wavelength to 1200 nm
and the stop wavelength to 1650 nm.
To limit the wavelength range
1
Press the Setup key.
2
Press the
WL LIM
softkey.
3
Press the
LIM ON
softkey if it is not already highlighted.
4
Press the
STARTWL
softkey to adjust the start wavelength value.
5
Press the
STOP WL
softkey to adjust the stop wavelength value.
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......