Using the Multi-Wavelength Meter
Displaying Wavelength and Power
Peak WL mode
When Peak WL is pressed, the display shows the largest amplitude line in the spec-
trum. This is the peak wavelength mode. The word
PEAK
is shown on the screen. If
multiple laser lines are present at the input, the number of lines located will be
shown along the right side of the screen.
Display after “Peak WL” key pressed
In addition to the digital readouts, there is a power bar. It provides a convenient ana-
log “meter movement” for tuning laser power.
Although the Peak WL mode shows one signal at a time, softkeys are provided that
allow you to scroll through and display all the measured laser lines. You can scroll
through the list according to the wavelengths or powers measured. The signals are
displayed in order from shortest to longest wavelengths. The Agilent 86120B can
measure up to 100 laser lines simultaneously.
To display peak wavelength and power
1
Connect the fiber-optic cable to the front-panel OPTICAL INPUT connector.
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......