165
Programming Commands
CALCulate2 Subsystem
POINts?
Queries the number of points in the data set.
Syntax
:CALCulate2:POINts?
Attribute Summary
Preset State: unaffected
*RST State: unaffected
SCPI Compliance: instrument specific
Query Only
Description
This is the number of points that will be returned by the CALC2:DATA? query.
Query Response
For example, if six laser lines are located:
+6
PTHReshold
Sets the peak threshold limit used by the instrument to determine valid laser line
peaks.
Syntax
:CALCulate2:PTHReshold{?| {<integer> | MINimum | MAXimum | DEFault}}
<integer>
represents logarithmic units in dB. Valid range is 0 to 40.
Constant
Value
MINimum
0 dB
MAXimum
40 dB
DEFault
10 dB
Attribute Summary
Non-sequential command
Preset State: 10 dB
*RST State: 10 dB
SCPI Compliance: instrument specific
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......