222
Performance Tests
Test 2. Sensitivity
Test 2. Sensitivity
Description
Sensitivity is verified using the following devices:
• Optical power meter
• Optical attenuator
• 1310 nm and 1550 nm lasers (>0 dBm output power)
C A U T I O N
Do not 18 dBm source power. The Keysight 86120B’s input
circuitry can be damaged when total input power exceeds 18 dBm.
Procedure
Perform the following procedure first using the 1310 nm laser and then repeat the
steps using the 1550 nm laser.
1
Connect the laser’s output to the optical attenuator’s input.
2
Connect the optical attenuator’s output to the optical power meter.
3
Adjust the attenuator for a reading of 0 dBm on the power meter. Record the
attenuator’s setting.
Attenuation at 0 dBm: ____________
4
Adjust the attenuator for a reading of –35 dBm on the power meter. Record the
attenuator’s setting.
Attenuation at –35 dBm: ____________
5
Disconnect the fiber-optic cable at the power meter’s input and connect the cable to
the Keysight 86120B being tested.
6
Reset the optical attenuator to the setting recorded in
7
Read the power and wavelength measured on the Keysight 86120B, and compared
them to the specifications listed in
Chapter 7, “Specifications and Regulatory
.
8
Reset the optical attenuator to the setting recorded in
9
Read the power and wavelength measured on the Keysight 86120B, and compared
them to the specifications listed in
Chapter 7, “Specifications and Regulatory
.
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......