Programming Commands
SENSe Subsystem
CORRection:ELEVation
Sets the elevation value used by the instrument to compensate for air dispersion.
Syntax
:SENSe:CORRection:ELEVation{?| {<integer> | MINimum | MAXimum}}
<integer>
is the altitude in meters.
Constant
Description
MINimum
0 m
MAXimum
5000 m
Attribute Summary
Non-sequential command
Preset State: unaffected by
*RST sets this value to the minimum.
SCPI Compliance: instrument specific
Description
Changing the elevation value causes the current data to be reprocessed.
Query Response
The query form returns the current elevation setting as shown in the following
example:
+1500
CORRection:MEDium
Sets the Keysight 86120B to return wavelength readings in a vacuum or standard
air.
Syntax
:SENSe:CORRection:MEDium{?| {AIR | VACuum}}
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command
has the time to complete before sending any more commands to the instrument.
Refer to “Always force the Keysight 86120B to wait for non-sequential com-
mands” on page 92
for more information.
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......