169
Programming Commands
CALCulate2 Subsystem
WLIMit:STARt[:WAVelength]
Syntax
CALCulate2:WLIMit:STARt[:WAVelength] {?|{<real> | MINimum | MAXimum}}
<
real
> is a wavenumber value that is within the following limits:
Constant
Description
MINimum
700.0 nm
MAXimum
wavelength limit stop value
Attribute Summary
Non-sequential command
Preset State: 700 nm
*RST State: 700 nm
SCPI Compliance: instrument specific
Description
This command sets the starting range for the wavelength limit. The start wavelength
value must be less than or equal to the stop wavelength value or the start wavelength
will be clipped to the stop wavelength and a “Data out of range” error will be
generated. Setting the start wavelength is equivalent to setting the stop frequency/
wavenumber because of the inverse relationship of frequency to wavelength. The
default units for the <
real
> parameter are meters.
Non-sequential command
Always use an *OPC? query or a *WAI command to ensure that this command
has the time to complete before sending any more commands to the instrument.
Refer to “Always force the Keysight 86120B to wait for non-sequential com-
mands” on page 92
for more information.
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......