175
Programming Commands
CALCulate3 Subsystem
:REFerence
:FREQuency
[:WAVelength]
:WNUMber
[:STATe]
ASNR:CLEar
Clears the number of measurements used in the average signal-to-noise calculation.
Syntax
:CALCulate3:ASNR:CLEar
Attribute Summary
Preset State: not affected
*RST State: not affected
SCPI Compliance: instrument specific
Description
This command clears the number of measurements used in the average signal-to-
noise calculation. The current measurement is used as the new reference for the
average signal-to-noise calculation.
ASNR:COUNt
Sets the number of measurements to be used for the average signal-to-noise calcula-
tion.
Syntax
:CALCulate3:ASNR:COUNt {?|{<integer> | MINimum | MAXimum }}
<integer> is a value that is within the following limits:
Constant
Description
MINimum
10
MAXimum
900
Attribute Summary
Preset State: 100
*RST State: 100
SCPI Compliance: instrument specific
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......