Step 5. Enter Your Elevation
Step 5. Enter Your Elevation
In order for your Keysight 86120B to accurately measure wavelengths and meet its
published specifications, you must enter the elevation where you will be performing
your measurements.
1
Press the Setup key.
2
Press the
MORE
softkey.
3
Press the
CAL
softkey.
4
Press
ELEV
.
5
Use the
and
softkeys to enter the elevation in meters. Entries jump in 500
meter steps from 0 m to 5000 m.
The elevation value selected with the softkeys must be within 250 meters of the ac-
tual elevation.
6
Press
RETURN
to complete the entry.
Converting feet to meters
If you know your elevation in feet, you can convert this value to meters by
using the following equation:
m
ft
3.281
-------------
=