154
Programming Commands
Measurement Instructions
MEASure{:ARRay | [:SCALar]} :POWer:WNUMber?
Returns a wave number value.
Syntax
:POWer:WNUMber? [<expected_value>[,<resolution>]]
Description
When used with a :SCALar command, a single value is returned. The display is
placed in the single-wavelength mode, and the marker is placed on the signal having
a wave number that is closest to the
<expected_value>
parameter. Default units for
<expected_value>
parameter are in m
–1
.
When used with an :ARRay command, an array of wave number is returned. The
display is placed in the list-by-wavelength mode.
The
<resolution>
parameter sets the resolution of the measurement. It is a unitless
number whose value will be limited to either 0.01 or 0.001 (whichever is closer).
Returned values are in inverse meters. Displayed units are inverse centimeters.
Power units are not affected.
<expected_value>
Constants
MAXimum
The laser line having the largest wave number
MINimum
The laser line having the smallest wave number
DEFault
The current marker position
<resolution>
Constants
MAXimum
0.01 resolution (fast update)
Used With
<expected_value>
<resolution>
SCALar
optional
optional
ARRay
ignored
a
a. Although ignored, this argument must be present if the resolution argument is specified.
optional
CONFigure command
When this function is used with the CONFigure command, the query question
mark character “?”
must
not
be included in the string. However, the FETCh,
READ, and MEASure command are queries and require the question mark.
Refer to the examples for this command.
Summary of Contents for 86120B
Page 1: ...Keysight 86120B Multi Wavelength Meter User s Guide...
Page 2: ......
Page 3: ...1...
Page 12: ...10 General Safety Considerations...
Page 68: ...You can use the ABORT and CONT softkey to stop and restart a print job that is in progress...
Page 100: ...98 Programming Monitoring the Instrument...
Page 231: ...229 Performance Tests Test 5 Amplitude Accuracy and Linearity...
Page 232: ......
Page 244: ...242 Specifications and Regulatory Information Product Overview...
Page 250: ...248 Reference Menu Maps Display List by Power Menu Display List by WL Menu...
Page 251: ...249 Reference Menu Maps Delta On Menu Delta Off Menu...
Page 253: ...251 Reference Menu Maps System Print Menu...
Page 254: ...252 Reference Menu Maps System Setup Menu...
Page 272: ......