LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 10 Care & Maintenance
10 - 13
10-6-3 Leakage Current Limits (cont’d)
Table 10-12 Equipment Type and Test Definitions
Applied Parts
(AP)
Parts or accessories that contact the patient to perform their function. For ultrasound
equipment, this includes transducers, ECG leads and e-TRAX Needle Sensor.
Type BF
Body Floating or non-conductive ultrasound
probes which are marked with the 'man in box'
BF symbol. this includes all transducers and
ECG leads
or
Type CF
Cardiac Floating or non-conductive,
intraoperative probes for direct cardiac contact,
isolated ECG connections and e-TRAX Needle
Sensor which are marked with the 'heart in box'
CF symbol.
or
Sink
Leakage
The current resulting from the application of mains voltage to the applied part. This test
is required test for Type CF applied parts.