Section 8-6 - Replacing Top Console Parts
8-6-15-5
Probe Cable Hooks Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
If all are successful, include the following debrief script: LOGIQ E9 Service Manual, Direction
5535208-100, Rev. 2. Equipment passed all required checks and is ready for use.
Table 8-108 Probe Cable Hooks replacement Functional Checks
See:
Section
Functional Check
Debrief Script
4-2-2
Power ON/Boot Up
LOGIQ E9 Service Manual, Direction
5535208-100, Rev. 2. Equipment passed all required
checks and is ready for use.
4-2-3
Power shut down
4-3-14
Probe/Connectors Checks
4-3-5
B-Mode Checks
4-3-7-3
Color Mode Checks
4-3-7-6
PW/CW Doppler Mode Checks
4-3-15
Peripherals Checks
4-2-4
Top Console position adjustment