Chapter 6 Service Adjustments
6 - 3
6-2-3
LCD Adjustment Procedure
6-2-3-1
Brightness
•
To reduce the brightness, press the left arrow button. A popup box with the current brightness
setting will be displayed on the screen.
•
To increase the brightness, press the right button.
•
Push the mode button once to remove the popup bow from the screen and save the current values.
6-2-3-2
Contrast
Use the recommended LCD settings. The contrast is adjusted by the LOGIQ E9 software, when
selecting scanning modes and doing scanning adjustments. Changing the monitor’s contrast settings in
one mode will influence the picture quality in other modes.
Table 6-1
Recommended LCD Settings
Room Condition
Contrast
Brightness
Dark Room for Radiology/Cardiology
85
35
Dim Room for Radiology/Cardiology
90
55
Bright Room
100
100
Factory Settings
100
60