GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
8 - 188
Section 8-7 - Main Console parts replacement
2.
Flip the Fan Tray over on a clean, safe
surface.
Pull out the Filter latch and slide the Filter
out.
Clean the Bottom Filter after it is removed
by removing excess lint or dust from the
soiled side. If necessary, use a clean, soft
brush; or vacuum.
3.
Reinstall the Filter into the Fan Tray and reinstall the Fan Tray.
Make sure the Fan Tray seats
completely into the Card Cage.
Table 8-114 Removing the Fan Tray Filter
Steps
Corresponding Graphic
Summary of Contents for 5205000
Page 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Page 2: ......
Page 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Page 807: ......
Page 808: ......