GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
7 - 38
Section 7-5 - Common Service Desktop
7-5-6-8
Touch Screen Tab
Test the Touch Screen functionality by pressing on the Touch Screen of the upper operator panel. Verify
that the press is indicated in the black window by a highlighted circle. See:
Figure 7-45
or
Figure 7-46
.
Figure 7-45 Touch Screen Tab - R4 and earlier
Figure 7-46 Touch Screen Tab - R5 and later
Check calibration
:
Touch screen and verify dots follow/
Calibrate: Press and hold each touch
target until it moves to next target.
Select Touch screen test, press each
target as they appear/Calibrate: Press
Run Windows calibration. Press very
lightly on each target until it moves to
next location.
Summary of Contents for 5205000
Page 1: ...10 20 14 GEHC_FRNT_CVR FM LOGIQ E9 SERVICE MANUAL Part Number 5535208 100 Revision Rev 2 ...
Page 2: ......
Page 21: ...GE DIRECTION 5535208 100 REV 2 LOGIQ E9 SERVICE MANUAL 19 ZH CN KO ...
Page 807: ......
Page 808: ......