Chapter 8 Replacement procedures
8 - 165
8-6-11-6
Calibration and adjustments
No calibrations or adjustments are needed after this part replacement.
8-6-11-7
Verification
Perform the following steps to verify that the product is functioning as intended after this replacement:
1.) Verify that all screws removed earlier have been installed.
2.) If finished, connect cables and Probes removed earlier.
3.) Power up the system to verify that it operates as intended.
4.) If a Gel Warmer is installed in the Options Holder, see: additional Verification and Functional
Checks for Gel Warmer (see:
8-6-12 "Gel Warmer replacement" on page 8-166
).
8-6-11-8
Functional Checks
Perform the following functional checks to confirm the system is operational before returning the system
to the customer.
If all are successful, include the following debrief script: LOGIQ E9 Service Manual, Direction
5535208-100
, Rev. 2. Equipment passed all required checks and is ready for use.
Table 8-106 Options Holder / Left or Right Support replacement Functional Checks
See:
Section
Functional Check
Debrief Script
4-2-2
Power ON/Boot Up
LOGIQ E9 Service Manual, Direction
5535208-100, Rev. 2. Equipment passed all
required checks and is ready for use.
4-2-3
Power shut down
4-3-14
Probe/Connectors Checks
4-3-5
B-Mode Checks
4-3-7
System CFM and PWD Checks
4-3-10
Basic Measurements
4-3-18
Mechanical Function Checks
4-3-18-2
Brakes and Direction Lock Checks
Summary of Contents for 5205000