Chapter 3 LOGIQ E9 Setup
3 - 71
3-9-4-9 How to Setup and Use a DICOM Query/Retrieve (Q/R) Service (cont’d)
Q/R Support for Volume Navigation
V Nav (Volume Navigation) combines 3D images from other modalities (CT, MR, etc.) with Ultrasound
scanning. Use Q/R to load non-Ultrasound images into the .
The supports retrieval of these non-US image types:
•
MR (Magnetic Resonance)
•
CT (Computed Tomography)
•
PT (Positron Emission Tomography)
•
XA (X-Ray Angio)
•
MG (Digital Mammography)
Use filter “00080061 Modalities in Study” to retrieve non-US images. To filter on more than one modality,
separate each by a “\“ (backslash). For example, “US\MR\CT\PT\XA\MG” returns entries for all
modalities supported by the .
Basic Use with V Nav
This guide is not intended to cover V Nav extensively. However, here are the initial steps:
Table 3-27 Q/R Support for Volume Navigation
Image History Display
1.
Query and retrieve the non-US study from PACS.
2.
Create a new exam for this patient.
3.
Press the
V Nav
key located above the keyboard on the operator panel.
4.
On the Touch Panel, press
Load
; select
From Database
from the pop-up menu.
5.
An Image History page displays on the screen. Select any desired image in the non-US series
and press
Review
.
(continued on next page)