Section 8-10 - Main Power Supply parts replacement
8-10-4-4-4
4D MC Reinstallation for Lambda PS
Table 8-168 4D MC Reinstallation for Lambda PS
Steps
Corresponding Graphic
1.
Install the two screws (circled),
supplied with the replacement 4D MC
to secure the 4D MC to the 4D MC
Cover.
NOTE: The replacement 4D MC
includes a two USB Cables, the longer
USB Cable must be used for a Lambda
PS.
Connect the long USB Cable to the 4D
MC. Route the cable as shown.
2.
Insert the 4D MC and cover into the PS,
along the mounting rails (circled).
NOTE: There are four mounting rails,
two at the top (circled) and two in the
bottom. Make sure all are aligned to
ensure the 4D MC connects to the
Lambda PS. The 4D MC connects
directly to the Lambda PS.