LOGIQ E9 S
ERVICE
M
ANUAL
7 - 8
Section 7-3 - Gathering Troubleshooting Data
7-3-5 Capturing Network Logs with Network Sniffer (Software R1.x.x) (cont’d)
7.) On the Select Filter Task screen, select “Show only IP packets FROM/TO one or more IP
addresses”, then select Next.
The Enter Filter Settings screen opens.
8.) Enter the IP address you wish to monitor, and click the Add button (). Repeat this step for all IP
addresses you wish to add, then select Finish.
The Enter Filter Settings screen closes.
9.) On the Configuration Screen, select the Capture Tab (
Figure 7-8 "Capture Tab on Configuration
Screen" on page 7-9
).
Figure 7-7 Enter Filter Settings