Chapter 7 Diagnostics/Troubleshooting
7 - 75
7-5-10
Diagnostics - BEP
For Diagnostic differences, see
7-5-12 "Diagnostics - BEP Interactive Tests - R4.x and later" on page
7-83
.
7-5-10-1
Essential Tests
See:
7-5-3 "Diagnostics Window Overview" on page 7-22
for a description of this window’s features.
Runs tests on essential functions of the BEP. This is a good starting point for isolating issues that may
originate from the BEP. The test results are displayed in the Status portion of the window.
Run Time: About 2 minutes.
1.) Enter the desired Loop Count (numeric value of 4 digits or less).
2.) Click the Execute button.
3.) Failed tests may have another diagnostic test that can help isolate the failure, or will indicate which
part to replace. For:
-
PCI Bus configuration
-
CMOS memory
-
CPU - run
7-5-10-3 "System Board" on page 7-76
-
RAM - run
7-5-10-9 "Memory" on page 7-81
-
Hard Drive status and Random Seek - run
7-5-10-4 "Hard Disk Drive Surface Scan" on page
7-77
and run
7-5-10-5 "Hard Disk Drive Quick Test" on page 7-78
-
Network Interface (loopback only) - run
7-5-10-7 "Network Adapter" on page 7-80
.
7-5-10-2
Essential Tests differences in R4 and later
Essential test in R4 has two versions:
•
Essential test: Complete test, same as R3 or earlier
Essential test, No network: same test, except it does not run the network tests for wired or wireless
connection.