HDD removal
If the HDD has patient data, leave the HDD at the site.
Table 8-140 HDD removal
Steps
Corresponding Graphic
1.
Loosen the thumb screw or screws at the top of the BEP cover that attach the cover to the BEP.
Tilt the top of the BEP cover away from the BEP.
2.
Disconnect:
• the battery cable from
J3
on the EPS
(Extended Power Shutdown) Board.
• the cables (
1
) connecting the EPS to the
BEP.
Set the BEP cover aside.
inside of BEP Cover, EPS and cables
3.
Remove the screw that secures the HDD
bracket.