7-3-5 Capturing Network Logs with Network Sniffer (Software R1.x.x) (cont’d)
17.)On the Distinct Network Monitor screen (
Figure 7-4 "Distinct Network Monitor Window" on page 7-
6
), select Capture > Stop.
After stopping the data collection, a screen similar to
Figure 7-11 "Save As Capture File dialog box"
on page 7-11
opens
Figure 7-10 Capture Data Sample
18.)Select
File > Save As
.
Enter the file name “d:\log\Sniffer\MyLog“.
19.)Select Save.
20.)On the Save As Capture File dialog (Figure 7-11), select Save.
Figure 7-11 Save As Capture File dialog box
21.)Check the d:\log\Sniffer directory. At a minimum, there should be the following two files: MyLog.cap,
MyLog.num.
22.)Copy the files to media, or ask the OLC to upload the files if the system has remote connectivity.