GE
D
IRECTION
5535208-100, R
EV
. 2
LOGIQ E9 S
ERVICE
M
ANUAL
Chapter 3 LOGIQ E9 Setup
3 - 15
3-5-3-1 Back Cover Label and ETL testing laboratories safety rating label (ETL may not be present)
(cont’d)
3-5-4
EMI protection
This unit has been designed to minimize the effects of Electro-Magnetic Interference (EMI). Many of the
covers, shields, and screws are provided primarily to protect the LOGIQ E9 from image artifacts caused
by this interference. For this reason, it is imperative that all covers and hardware are installed and
secured before the unit is put into operation.
See:
2-2-3 "EMI limitations" on page 2-5
for more information about EMI protection.
Figure 3-6 Rating Plate Label - R4.x.x and Later