Chapter 3 Working with a TI Calculator
123
4. Press
to begin calibrating
MultiLab displays the current slope and offset. You should enter two
actual values and the corresponding read values (the read values are
the values displayed by MultiLab when measuring the two actual
values)
5. Enter the 1
st
actual value, then press
6. Enter the 1
st
read value, then press
7. Repeat the last two steps for the other value
MultiLab will calculate the new slope and offset and will return to the
setup screen
2. To Restore the Default Calibration Values
1. Navigate to the CALIBRATION OPTIONS screen (see above)
2. Press
MultiLab will restore the default slope and offset and will return to the
setup screen