7 Connection
7.1 Transducers
FLUXUS G831ST-HT
2021-06-01, UMFLUXUS_G831ST-HTV1-0EN
58
7.1.5
Transducer module (SENSPROM)
The transducer module contains important transducer data for the operation of the transmitter with transducers.
If transducers are replaced or added, the transducer module has to be replaced or added as well.
• Disconnect the transmitter from the power supply.
• Open the front side of the upper housing.
• Remove the module.
• Insert the transducer modules into their sockets.
• Push the module into the housing.
• Close the housing.
• Make sure that the housing is closed correctly and that the set screws are tightened.
• Connect the transmitter to the power supply.
• Enter all parameters of the program branch
Parameters
.
• Start the measurement.
Notice!
The serial numbers of the transducer module and the transducer have to be identical. A wrong or incorrectly
connected transducer module will lead to incorrect measured values or to a measurement failure.
Fig. 7.8:
Connection of the transducer module
1 – transducer module (measuring channel A)
2 – transducer module (measuring channel B)
2
1
2
1