14 Inputs
14.2 Assignment of an input
FLUXUS G831ST-HT
2021-06-01, UMFLUXUS_G831ST-HTV1-0EN
114
14.1.4
Terminal assignment
The terminals for the connection of the input are displayed.
By pressing
further information is displayed.
• Press ENTER.
14.1.5
Function test of the input
The function of the installed input can now be tested.
Analog input
• Connect the signal source to the input.
• Select
Yes
to test the input signal. Select
No
to display the next menu item.
• Press ENTER.
• If the transmitter displays a value (here:
Current
), the input functions correctly.
• Press ENTER.
• Select
Repeat
to repeat the test or
Finish
to display the next menu item.
• Press ENTER.
• Select
Yes
to test the assignment of the measured value to the input signal. Select
No
to display the next menu item.
• Press ENTER.
• If the transmitter displays a value (here:
Temperature
), the input functions correctly.
• Press ENTER.
• Select
Repeat
to repeat the test or
Finish
to display the next menu item.
• Press ENTER.
• Press BRK to return to the program branch
Options
.
14.2
Assignment of an input
• Select the menu item
Assign inputs
.
• Press ENTER.
• Select a list item. Only the installed inputs are displayed in the scroll list.
• Select the list item
No linkage
if no input is to be assigned.
• Press ENTER.
Inputs\...\Input info
Special functions\Inputs\...\I1 Test signal
Special functions\Inputs\...\I1 Test signal\Current
Special functions\Inputs\...\I1 Test mea. range
Special functions\Inputs\...\I1 Test mea. range\Temperature
Options\Assign inputs