11 Maintenance and cleaning
FLUXUS G831ST-HT
11.1 Maintenance
99
UMFLUXUS_G831ST-HTV1-0EN, 2021-06-01
11.1
Maintenance
The transmitter and the transducers are practically maintenance-free. In order to ensure security, the following
maintenance intervals are recommended:
11.2
Cleaning
Stainless steel housing
• Clean the housing with a soft cloth and care and cleaning spray for stainless steel.
Aluminum housing
• Clean the housing with a soft cloth. Do not use detergents.
Transducers
• Remove traces of coupling compound from the transducers with a soft paper towel.
11.3
Calibration
If installed as recommended in an appropriate location, used cautiously and serviced conscientiously, no troubles should
appear.
The transmitter has been calibrated at factory and, usually, a recalibration is not necessary.
A recalibration is recommended if:
• the contact surfaces of the transducers show visible wear
• the transducers were used for a prolonged period at high temperatures (several months > 130 °C for normal transducers
or > 200 °C for high temperature transducers)
In order to realize a recalibration under reference conditions, either the transmitter, the transducers or both have to be sent
to FLEXIM.
item
maintenance step
interval
measure
stainless steel housing
• junction box
• transducer mounting fixture
visual inspection for corrosion
and damages
annually
cleaning
visual inspection for
contamination
annually or more frequently,
depending on the ambient
conditions
aluminum housing
• transmitter
visual inspection for
contamination
annually or more frequently,
depending on the ambient
conditions
transducers
check of the transducer
coupling on the pipe
annually
replacement of coupling foil, if
necessary
O-rings
visual inspection for fissures
annually
see section 11.4
transmitter
check for firmware updates
annually
update, if necessary
transmitter
functional test
annually
reading of measured and
diagnostic values
transmitter and transducers
calibration
-
see section 11.3