6 Mounting
FLUXUS G831ST-HT
6.2 Transducers
31
UMFLUXUS_G831ST-HTV1-0EN, 2021-06-01
Step 2: Fixation of the clasps to the tension straps
• Select the installation instruction of the supplied clasp:
Band clamp clasp
The clasp is fixed to the tension strap, see Fig. 6.8.
Quick release clasp
The clasp is fixed to the tension strap, see Fig. 6.9.
• Cut the tension strap to length (pipe circumf at least 120 mm).
Ratchet clasp
• Cut the tension strap to length (pipe circumf at least 120 mm).
• Insert approx. 100 mm of the tension strap into part (1) and (2) of the ratchet clasp, see Fig. 6.10 a.
• Bend the tension strap.
• Insert the tension strap into part (1) of the ratchet clasp, see Fig. 6.10 b.
• Tighten the tension strap.
• Repeat the steps for the second tension strap.
Fig. 6.8:
Band clamp clasp with tension strap
Fig. 6.9:
Quick release clasp with tension
strap
Caution!
The edge of the tension strap is very sharp.
Risk of injury!
→ Debur sharp edges.
→ Wear the required personal protective equipment.
→ Observe the applicable rules.
Fig. 6.10: Ratchet clasp with tension strap
a
b
1
2
1