17 Advanced functions
17.4 Remote functions
FLUXUS G831ST-HT
2021-06-01, UMFLUXUS_G831ST-HTV1-0EN
126
17.4
Remote functions
Remote functions can be triggered by event triggers,
triggerable analog inputs or binary inputs.
In order to define an input for a remote function, it has to be
enabled in the menu item
Special functions\Inputs
.
In order to define an event trigger for a remote function, it
has to be enabled in the menu item
Options\Event
trigger
.
It is possible to trigger one or more of the following remote functions:
• reset of measured values
• reset of totalizers
• stop of totalizers
17.4.1
Configuration of the remote function
• Select the menu item
Options\Remote functions
.
• Press ENTER.
In the scroll list of the functions it is displayed whether, and if so, which input or event trigger is assigned to a function.
• Select a list item:
– Reset meas. val. (-)
– Reset totalizers (-)
– Stop totalizers (-)
If an input or event trigger has already been assigned to this function, this is displayed as follows:
Reset meas.
val. (R1)
.
Reset of measured values
• Select the list item
Reset meas. val.
• Press ENTER.
The measured value output simulates a reposing application for the duration of the signal. The actual measured flow
velocity is ignored and the measured value is set to zero. All values of the physical quantity derived from the flow velocity
and the values from the calculation channels also yield zero.
The transmitter continues the measurement if the condition of the remote function is no longer met.
• Select the input which is to be used to trigger the selected remote function.
• Press ENTER.
• Select
No linkage
in order to deactivate the remote function.
• Press ENTER.
Options
Remote functions
→ Reset meas. val.
→ Reset totalizers
→ Stop totalizers
With input
see annex A
Options\Remote functions
Notice!
The settings made will be stored at the beginning of a measurement.