17 Advanced functions
17.1 Totalizers
FLUXUS G831ST-HT
2021-06-01, UMFLUXUS_G831ST-HTV1-0EN
122
The number of decimal places is constant. The max. value of the totalizers decreases with an increasing number of
decimal places.
17.1.2
Detection of long measurement failures
If there are no valid measured values during a long time interval, the totalizers remain unchanged.
The time interval can be defined.
• Select the menu item
Totalizers
in the program branch
Special functions
.
• Press ENTER until the menu item
Totalizer timeout
is displayed.
• Select
Default
if no customized inputs are to be carried out and the default value of 30 s is to be used.
• Press ENTER.
• Select
Customized
if the time interval is to be defined.
• Press ENTER.
• Enter the time interval.
• Press ENTER.
17.1.3
Totalizer overflow
The overflow behavior of the totalizers can be set:
Without overflow
• The value of the totalizer increases to the internal limit of 10
38
.
• The values will be displayed as exponential numbers (±1.00000E10), if necessary. The totalizer can only be reset to zero
manually.
With overflow
The totalizer will be reset to zero automatically when ±9999999999 is reached.
• Select the menu item
Totalizers
in the program branch
Special functions
.
• Press ENTER until the menu item
Overflow behavior
is displayed.
• Select
Yes
to work with overflow. Select
No
to work without overflow.
• Press ENTER.
Independent of the setting, the totalizers can be reset to zero manually.
decimal places
max. value
max. display
0
< 10
10
±9999999999
1
< 10
8
±99999999.9
2
< 10
7
±9999999.99
3
< 10
6
±999999.999
4
< 10
5
±99999.9999
Notice!
The number of decimal places and the max. value of the totalizers only affect the display.
Special functions\Totalizers\Totalizer timeout
Special functions\Totalizers\Overflow behavior