9 Measurement
FLUXUS G831ST-HT
9.2 Measurement settings
81
UMFLUXUS_G831ST-HTV1-0EN, 2021-06-01
9.2.4
Input of the error delay
The error delay is the time interval after which the error value is transmitted to the output.
This display will only be indicated if the list item
Edit
is selected in the menu item
Special functions\Dialogs/
Menus\Error value delay
.
If no error delay is entered, the damping factor will be used.
• Enter a value for the error delay.
• Press ENTER.
• Press BRK to return to the program branch
Options
.
9.2.5
Configuration of an output
If a current output has to be operated according to NAMUR NE43, this function has to be enabled.
• Select the list item
Current output
in the menu item
Special functions
.
• Press ENTER until the menu item
NAMUR NE43
is displayed.
• Select
Yes
to enable NAMUR NE43.
• Press ENTER.
The measured value, the status value or an event value
can be transmitted via the different outputs.
In the following, the configuration of an analog output is
described.
• Select the program branch
Options
.
• Press ENTER.
Assignment of an output
• Select the menu item
Outputs
.
• Press ENTER.
• Select the output (here:
Current I1(--)
).
• Press ENTER.
The scroll list contains all available outputs of the transmitter:
If the output has already been enabled, it is displayed as follows:
I1 (
)
.
• Select
Yes
to change the settings for an already enabled output.
• Select
No
to cancel the assignment and to return to the previous menu item.
• Press ENTER.
Options\Measurement\Error value delay
Special functions\Current output\NAMUR NE43
Options
Outputs
Select output
Enable output
Source item
see annex A
Options\Outputs\Current I1(--)
Options\Outputs\Current I1\Enable I1