7 Connection
7.1 Transducers
FLUXUS G831ST-HT
2021-06-01, UMFLUXUS_G831ST-HTV1-0EN
52
7.1.1
Connection of the transducer cable to the transmitter
• Remove the blind plug for the connection of the transducer cable.
• Insert the transducer cable into the housing.
• Fix the transducer cable by tightening the cable gland.
• Connect the transducer cable to the terminals of the transmitter.
Important!
The degree of protection of the transmitter is only ensured if all cables are tightly fitted using cable glands and the
housing is firmly screwed.
Tab. 7.1:
Terminal assignment
terminal
connection
AV
transducer (brown cable, marked white)
AVS
transducer (red cable)
ARS
transducer (red cable)
AR
transducer (brown cable)
Fig. 7.3:
Connection of the transducer cable with stainless steel conduit and stripped ends to the
transmitter
1 – cap nut
2 – compression part
3 – basic part
AR
AV
ARS AVS
BR
BV
BRSBVS
L
(+)
N
(-)
12
11
10
9
8
7
6
5
4
3
2
1
cable gland
2
3
1