13 User modes
13.1 StandardUser mode
FLUXUS G831ST-HT
2021-06-01, UMFLUXUS_G831ST-HTV1-0EN
106
13.1.5
High voltage
High voltage increases the excitation voltage of the transducers. When measuring using the correlation principle, high
voltage is activated by default.
• Select the menu item
Special settings
in the program branch
Options
.
• Press ENTER until the menu item
High voltage
is displayed.
• Select
Off
to deactivate high voltage.
• Press ENTER.
13.1.6
Averaging factor of the measurement
The averaging factor of the measurement defines a time range in which the values are averaged for calculation. A high
factor increases the measurement stability, but also the response time to changed measured values.
• Select the menu item
Options
in the program branch
Special settings
.
• Press ENTER until the menu item
Averaging factor
is displayed.
• Enter the averaging factor (default: 5). The value should be between 3 and 7.
• Press ENTER.
Options\Special settings\High voltage
Options\Special settings\Averaging factor