7 Connection
7.5 Temperature probe
FLUXUS G831ST-HT
2021-06-01, UMFLUXUS_G831ST-HTV1-0EN
66
Connection of the extension cable to the junction box
• Remove the blind plug for the connection of the extension cable.
• Open the cable gland of the extension cable. The compression part remains in the cap nut.
• Push the extension cable through the cap nut and the compression part.
• Prepare the extension cable.
• Shorten the external shield and brush it back over the compression part.
• Screw the sealing ring side of the reducer into the junction box.
• Screw the basic part into the reducer.
• Insert the extension cable into the junction box.
• Fix the cable gland by screwing the cap nut onto the basic part.
• Connect the extension cable to the terminals of the junction box.
Fig. 7.15: Preparation of the extension cable
1 – cap nut
2 – compression part
3 – basic part
4 – reducer
5 – external shield, brushed back
6 – sealing ring side
Fig. 7.16: Junction box
1 – connection of the extension cable
2 – connection of the temperature probe
110 mm
7 mm
1
2
3
4
6
5
1
2
KL2
KL1