17 Advanced functions
FLUXUS G831ST-HT
17.5 Event triggers
127
UMFLUXUS_G831ST-HTV1-0EN, 2021-06-01
Reset of the totalizers
• Select the list item
Reset totalizers
.
• Press ENTER.
The totalizers are set to zero.
Totalizing starts at zero again, as soon as the condition for the remote function is no longer met.
When the totalizers are reset to zero with the remote function, the character
H
is displayed next to the measured value
during the measurement.
• Select the input which is to be used to trigger the selected remote function.
• Press ENTER.
• Select
No linkage
in order to deactivate the remote function.
• Press ENTER.
Stop of the totalizers
• Select the list item
Stop totalizers
.
• Press ENTER.
The totalizers are stopped for the duration of the signal.
Totalizing will be continued with the last registered totalizer value if the condition of the remote function is no longer met.
• Select the input which is to be used to trigger the selected remote function.
• Press ENTER.
• Select
No linkage
in order to deactivate the remote function.
• Press ENTER.
17.5
Event triggers
It is possible to configure max. 4 independent event
triggers per channel: R1, R2, R3 and R4.
The event triggers can be used, e.g., to:
– output information about the running measurement
– trigger special remote functions
– switch on/off pumps and motors
• Select the program branch
Options
.
• Press ENTER.
• Select the event trigger.
If an event trigger has already been installed, it is displayed as follows:
Rx(+)
.
• Select
Yes
to change the settings for an already assigned event trigger or to assign a new one.
• Select
No
to cancel the assignment and to return to the previous menu item.
• Press ENTER.
• Select the source item (physical quantity) for which a condition has to be defined.
Options
Event trigger
Select Rx
Enable Rx
Source item
Define the function
see annex A
Options\Event trigger\Rx(-)
Options\Event trigger\Enable Rx
Options\Event trigger\Enable Rx\Source item