9 Measurement
9.1 Parameter input
FLUXUS G831ST-HT
2021-06-01, UMFLUXUS_G831ST-HTV1-0EN
76
The parameters can also be entered without connected transducers or transducer module.
• Select the list item
Connected transd.
• Press ENTER.
The display
Transducer not found
will be indicated if neither transducers nor transducer module are connected.
• Press ENTER.
• Select the list item
Select transducer
to use standard transducers stored in the transmitter.
• Select the transducer.
• Press ENTER.
This display will not be indicated if the transducers and the transducer module are connected to the transmitter.
9.1.2
Input of pipe parameters
Outer pipe diameter
• Enter the outer pipe diameter.
• Press ENTER.
It is possible to enter the pipe circumference instead of the outer pipe diameter.
Pipe circumference
• Activate the input of the pipe circumference in the menu item
Special functions\Dialogs/Menus\Pipe
circumference
.
• Enter zero in the menu item
Outer diameter
. The menu item
Pipe circumference
will be displayed.
• Enter the pipe circumference.
• Press ENTER.
If the outer pipe diameter is to be entered, enter zero. The menu item
Outer diameter
is displayed.
Pipe material
The pipe material has to be selected to be able to determine the corresponding sound speed.
The sound speeds for the materials in the scroll list are stored in the transmitter.
• Select the pipe material.
• If the material is not in the scroll list, select the list item
Other material
.
• Press ENTER.
Parameters\Connected transd.
Parameters\Select transducer
Parameters\Outer diameter
Parameters\Pipe circumference
Parameters\Pipe material