6 Mounting
6.2 Transducers
FLUXUS G831ST-HT
2021-06-01, UMFLUXUS_G831ST-HTV1-0EN
36
Ratchet clasp
• Insert the tension strap into the tension strap clamp (2) and the metal spring, see Fig. 6.19 and Fig. 6.22.
• Place the tension strap around the pipe and insert it into the tension strap clamp on the opposite side and the slot of the
clasp screw (see Fig. 6.23).
• Position the metal spring opposite the ratchet clasp.
• Tighten the tension strap.
• Cut off the protruding tension strap, see Fig. 6.23.
• Tighten the clasp screw.
• Tighten the nuts of the tension strap clamps (2), but not too firmly in order not to damage the tension strap (see Fig.
• Repeat the steps for fixing the second pair of rails. When positioning the second pair of rails, observe the required
distance between the measuring barriers.
Caution!
The edge of the tension strap is very sharp.
Risk of injury!
→ Debur sharp edges.
→ Wear the required personal protective equipment.
→ Observe the applicable rules.
Notice!
In order to release the screw and the tension strap, press the lever down, see Fig. 6.16.
Fig. 6.22: Tension strap with metal spring and clamp
1 – metal spring
2 – tension strap clamp
Fig. 6.23: Ratchet clasp with tension strap
1 – sense of rotation
2 – edge
3 – lever
4 – clasp screw with slot
1
2
1
2
4
3
4