17 Advanced functions
FLUXUS G831ST-HT
17.1 Totalizers
123
UMFLUXUS_G831ST-HTV1-0EN, 2021-06-01
17.1.4
Totalizer behavior after the measurement is stopped
It is possible to define the totalizer behavior after the measurement is stopped or after a reset of the transmitter.
• Select the menu item
Totalizers
in the program branch
Special functions
.
• Press ENTER until the menu item
Keep totalizers
is displayed.
• Select
Yes
if the values of the totalizers are to be stored and used for the next measurement. Select
No
if the totalizers
are to be set to zero.
• Press ENTER.
17.1.5
Totalizer sum
The sum of the totalizers for both flow directions can be displayed in the status line during the measurement.
• Select the menu item
Totalizers
in the program branch
Special functions
.
• Press ENTER until the menu item
Show ΣQ
is displayed.
• Select
Yes
to display the totalizer sum. Select
No
if it is not to be displayed.
• Press ENTER.
17.1.6
Totalizer storing
The totalizer values can be stored now.
• Select the menu item
Special functions\Data logger\Configuration
.
• Press ENTER until the menu item
Store totalizers
is displayed.
• Select
Yes
.
• Press ENTER.
Notice!
The overflow of a totalizer influences all output channels, e.g., the data logger and the online transmission of data.
The sum of both totalizers (throughput
Q
) transmitted via an output is no longer valid after one of the totalizers has
overflowed for the first time.
Special functions\Totalizers\Keep totalizers
Special functions\Totalizers\Show ΣQ