9 Measurement
FLUXUS G831ST-HT
9.2 Measurement settings
87
UMFLUXUS_G831ST-HTV1-0EN, 2021-06-01
• Enter a test value. It has to be within the output range.
• Press ENTER.
If the external measuring instrument displays the entered value, the output functions correctly.
• Select
Repeat
to repeat the test or
Finish
to display the next menu item.
• Press ENTER.
• Select
Yes
to test the status of the output signal. Select
No
to display the next menu item.
• Press ENTER.
• Select in the scroll list
Status OK
or
Status error
.
• Press ENTER.
If the external measuring instrument displays the value (min. output value for
Status error
, max output for
Status
OK
), the output functions correctly.
• Select
Repeat
to repeat the test or
Finish
to display the next menu item.
• Press ENTER.
9.2.5.3
Output of an event value
Output range
• Select a list item.
– 4...20 mA
– Other range
• Press ENTER.
If
Other range
is selected, enter the values
Output MIN
and
Output MAX
.
The output range has to be > 10 % of the max. output value (
Output MAX
). An error message will be displayed if the
output range is smaller. The next possible value will be displayed.
Idle state
The status of the output signal is defined which is to be output if no event occurs.
• Select the value for the idle state.
• Press ENTER.
Terminal assignment
The terminals for the connection of the output are displayed.
By pressing
or
further information is displayed.
• Press ENTER.
Options\Outputs\...\Enter test value
Options\Outputs\...\Test mea. range
Options\Outputs\...\Enter test value
Options\Outputs\...\Output range
Options\Outputs\...\Idle state
Options\Outputs\...\Output info