Table 3-12: Terminal Block Connections for the Rosemount 5900S with Single
Tankbus
Connection
Two-in-One / Single tankbus
X1: Primary Tankbus in Intrinsically safe Tankbus input, power and communication
X2: Primary Terminate
on
Termination for primary tankbus. The integrated line terminator is
connected over the Tankbus when a jumper is placed in the terminal
block.
X3: Primary Tankbus
out
Jumpers between X3 and X4
X4: Secondary Tankbus
in
Test terminals
Test terminals for temporary connection of a field communicator
Gauge terminal block SIL safety system
The Rosemount 5900S has a SIL3 alarm output which is connected to the Rosemount
2410 Tank Hub.
Figure 3-29: Terminal Compartment
'
&
$
%
)%
)%
A.
Test terminals
B.
2-in-1: optional jumpers between X3 and X4 for connection to second level gauge
C.
SIL Alarm
D.
Ground terminals, internal
FB.
Field Bus
For the 2-in-1 SIL version of the Rosemount 5900S, optional jumpers are placed between
terminals X3 and X4 for connection to the second radar level gauge.
Reference Manual
Installation
00809-0100-5900
September 2019
Rosemount 5900S Radar Level Gauge
89
Summary of Contents for Rosemount 5900S Series
Page 1: ...Reference Manual 00809 0100 5900 Rev EA September 2019 Rosemount 5900S Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5900 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5900 12 Reference Manual ...
Page 182: ...Configuration Reference Manual September 2019 00809 0100 5900 182 Reference Manual ...
Page 186: ...Operation Reference Manual September 2019 00809 0100 5900 186 Reference Manual ...
Page 266: ...Product Certifications Reference Manual September 2019 00809 0100 5900 266 Reference Manual ...