A. Gauge Reference Point
B. Distance from gauge reference point to Verification Pin
C. Verification Pin
Note
The value entered in the
Nominal Pos
field refers to the mechanical distance from
the Gauge Reference Point to the actual Verification Pin. This value will only act as a
starting point for the verification process in which the electrical distance from the
Gauge Reference Point to the Verification Pin is calculated. In most cases the
electrical distance deviates from the actual mechanical distance.
6. Ensure that the Threshold value is 500 mV.
The amplitude of the echo from the Verification Pin must be above the threshold
value in order to appear in the
LPGVerify
). In case the Verification Pin does not show up, it is possible to use a
smaller threshold value. Check that the product level is not above the Verification
Pin.
Reference Manual
Configuration
00809-0100-5900
September 2019
Rosemount 5900S Radar Level Gauge
125