4. In the
Registers Type
list, select
Predefined
or
All
.
Option
Description
Predefined
View a basic selection of registers.
All
View a range of registers by your own choice (for
advanced service).
5. For the All option, you have to specify a range of registers by setting a start value in
the
Start Register input
field, and the total number of registers to be displayed in
the
Number of Registers
field (1-500). Up to 50 registers is recommended for a
quick update of the list.
6. The
Registers Scope
drop-down list has three options:
Scope
Description
Access level
Basic
Standard setting that includes the most
commonly used registers
View Only
Service
Includes a wider range of registers for
advanced service and troubleshooting
Supervisor
Developer
For advanced users only
Administrator
7. In the
Show Values in
pane, choose the appropriate register format Decimal or
Hexadecimal.
8. Click the
Read
button.
Now the Value column is updated with the current register values.
Reference Manual
Service and troubleshooting
00809-0100-5900
September 2019
Rosemount 5900S Radar Level Gauge
189
Summary of Contents for Rosemount 5900S Series
Page 1: ...Reference Manual 00809 0100 5900 Rev EA September 2019 Rosemount 5900S Radar Level Gauge ...
Page 6: ...Contents Reference Manual September 2019 00809 0100 5900 vi Reference Manual ...
Page 12: ...Introduction Reference Manual September 2019 00809 0100 5900 12 Reference Manual ...
Page 182: ...Configuration Reference Manual September 2019 00809 0100 5900 182 Reference Manual ...
Page 186: ...Operation Reference Manual September 2019 00809 0100 5900 186 Reference Manual ...
Page 266: ...Product Certifications Reference Manual September 2019 00809 0100 5900 266 Reference Manual ...